Video: Isozentrumskontrolle of linear accelerators (radiation therapy)

method:
  1. A imaging plate (from the tape of the simulator) was positioned between two Plexiglas cylinder (d = 30cm) in the isocenter of the accelerator.
  2. fields, etc. were among the Tragarmwinkeln 0 °, 72 °, 144 °.
  3. The image information of the image plate was read into the irradiated Dicomarchiv.
evaluation of Dicomdatei
  • of the programmed software I DicoRView
  • The viewer is freeware (for use at your own risk).
  • takes about 4 minutes - the process is in the video to see here
  • if not - direct link


A feature of the Image plate helps in determining the Isozentrumsachsen:
light clears the slide> the positioning laser are clearly visible as white lines